The LED-R? Series of Radiometers from EIT comprises a new family of radiometers designed specifically to measure the energy generated by UV LED systems. The optics and instrument response have been engineered specifically for UV LEDs. The first product released in the LED-R Series is the EIT LEDCure? featuring a L395 response with a near rectangular response from 370-420 nm.
——The LED-R? Series of Radiometers from EIT comprises a new family of radiometers designed specifically to measure the energy generated by UV LED systems. The optics and instrument response have been engineered specifically for UV LEDs. The first product released in the LED-R Series is the EIT LEDCure? featuring a L395 response with a near rectangular response from 370-420 nm.
——美國EIT LED能量計
L395 Spectral Response | 370-420 nm (FWHM, 50 nm; Tolerance ± 2 nm); Out of Band, 4 OD Blocking |
L395 Dynamic Range | 40 W/cm2 |
Operating Range | 400 mW/cm2 to 40 W/cm2 Note: Operating Range provides best performance. Units will “turn on” and display data at much lower irradiance values |
Spatial Response | Approximately cosine |
Accuracy | ±10% of reading plus ±0.2% of full scale Typical ±5% or better |
Repeatability | ± 1-3% typical; dependent on source and equipment (conveyor) stability, unit alone better than 1% |
Smooth Modes | Smooth ON: Effective Sample rate of 25 samples/second Smooth OFF: Effective Sample rate of 2048 samples/second Smooth PROFILER: Effective Sample rate of 128 samples/second |
Sample Rate for Profiling | The LEDCure Profiler uses a fixed sample rate of 128 samples/second for profiling. For best matching between instrument display and PowerView Software? II values, use Smooth PROFILER mode |
Memory Capacity For Profiling | The memory capacity of the LEDCure in Profiler Mode is sufficient to collect data for >100 minutes |
PowerView SoftwareR II | National Instruments LabVIEW based programming designed for Windows XP, Windows NT, Windows Vista and Windows 7-10. Collected data stored in LabVIEW based *.tdms files |
Display | Easy to Read, Yellow Text on Black Background, Display Brightness User Adjustable |
Operating Temperature | 0-75°C Internal temperature; withstands high external temperatures for short periods (audible alarm indicates when temperature has exceeded upper limit ) |
Time-Out Period | 2 minutes DISPLAY mode (no key activity) |
Battery/Battery Life | 2 user-replaceable AAA Alkaline Cells, Approximately 20 hours with display on |
Instrument Dimensions and Materials | 4.60 x 0.50 inches; 117 mm x 12.7 mm (D x H), Aluminum & Stainless Steel |
Instrument Weight | 10.1 ounces (289 grams) |
Carrying Case | Supplied with carrying case, cut polyurethane foam interior, scuff resistant nylon exterior cover, Size: 10.75 x 3.5 x 7.75 inches; 274 x 89 x 197 mm (W x H x D), Weight: 9 ounces (260 grams) |
2018-05
The LED-R? Series of Radiometers from EIT comprises a new family of radiometers designed specifically to measure the energy generated by U…… [了解更多]